Nigen LCMS 40.0 / 40.1 / 100.0

FLOW RATE: 40 SLPM 40 SLPM 100 SLPM

PURITY: 98% to 99.9%

TECHNOLOGIES EMPLOYED:PSA

OUTPUT PRESSURE: 0 to 7 barg / 15 to 100 psig

OXYGEN AT MAX FLOW: < 2%

MOISTURE AT MAX FLOW: - 44°C dew-point STP, < 80 ppm

 

AIR INPUT CHARACTERISTICS: 

Particulate ≤ 0.01 microns

Oil vapors ≤ 0.01 mg / m3

Moisture ≤ 3°C dew point in pressure

Min pressure 8.5 barg / 120 psig

Max pressure 10 barg / 145 psig

AIR COMPRESSOR: No for NiGen LCMS 40.0 and  NiGen LCMS 100.0 - Yes for NiGen LCMS 40.1

 

ELECTRICAL SUPPLY: 

115 VAC ± 10%, 60 Hz

230 VAC ± 10%, 50 Hz

 

INDEX OF PROTECTIONIP20

OPERATING TEMPERATURE: 40°F to 100°F / 5°C to 40°C

ENVIRONMENTAL MOISTURE: ≤ 80%

MAX ALTITUDE: 2000 m or 6560 ft above sea level

 

DIMENSIONS: 

51.1”(H) x 17.3” (W) x 43.3” (D)

130 cm (H) x 44 cm (W) x 110 cm (D)

WEIGHT: 198 lbs / 90 kg 220 lbs / 100 kg 330 lbs / 150 kg

APPLICATIONS: LC-MS E, sample preparation


ADDITIONAL FEATURES

• Air + N2configuration for AB LC-MS

• High efficiency hydrocarbon removal columns included

• Air + N2configuration for Agilent Jet Stream LC-MS

• CPU for control of up to 32 units

• CAN-BUS communication

• Maintenance diary

• Report alarms

• Equipped with wheels


ACCESSORIES

• AB pressure regulator kit

• UHP available nitrogen purifier for Agilent collision cell on Jet Stream LC-MS

• Inlet air coalescing filter

• 3 point pressure regulator kit

• Installation kit

• Automatic 2 sources switch kit

DOCUMENTS 

Download Technical Features